Rohde & Schwarz RTC1002EDU oscilloscope

Rohde & Schwarz RTC1000 manual - Page 76

For RTC1002EDU.

PDF File Manual, 371 pages, Read Online | Download pdf file

RTC1002EDU photo
Loading ...
Loading ...
Loading ...
Analysis
R&S
®
RTC1000
76User Manual 1335.7352.02 ─ 04
3. To shift the mask vertically, or to enlarge it, press "Y-POSITION" and "STRETCH
Y", and use the Universal knob
4. To set the tolerance for the mask test, press "WIDTH Y" and "WIDTH X". Use the
Universal knob or the [KEYPAD] to enter values with a resolution of 1/100 division.
A mask has a minimum and a maximum value for each captured data value. The
minimum and maximum value for a source waveform with only one value per data
are identical. The width indicates the distance between the mask points and the
original point. The greater the selected value is, the greater are the potential wave-
form deviations in the amplitude.
The generated and edited mask can be used immediately for the test, however, it is
only saved temporarily in the instrument storage.
Masks are displayed as light white waveforms on the screen. If a mask was copied
or loaded, you can use the menu to change the limits for the test.
5. To store the mask permanently to a USB flash drive or in the internal memory,
press "SAVE".
6. To return to the "PASS/FAIL" menu, press the "Back" key.
To load a mask
1. In the "PASS/FAIL" menu, press "LOAD MASK" to open a file browser which allows
you to load previously stored masks for the test ).
2. In the file browser, select the mask file (file extension .MSK).
3. Load the selected file.
4. To change a loaded mask, access the "NEW MASK" menu.
A loaded mask is adjusted in the same way as a created mask. You can save the
changed mask.
To select an action:
1. In the "PASS/FAIL" menu, press "ACTIONS"
2. Select the actions:
Audio signal if the tolerance limits are exceeded
Stop at failure, the number of failures is adjustable
Pulse at the [Aux Out] output in case of failure
Screenshot at first failure, saved on USB flash drive
Print on connected printer
Waveform data saved to file
An action is performed if the condition is met, for example, a given number of mask
failures.
3. To return to the "PASS/FAIL" menu, press the "Back" key.
Pass/Fail Test Based on Masks
Loading ...
Loading ...
Loading ...