Remote Commands Reference
R&S
®
RTC1000
203User Manual 1335.7352.02 ─ 04
13.4.3 Width (Pulse) Trigger
TRIGger:A:WIDTh:POLarity............................................................................................ 203
TRIGger:A:WIDTh:RANGe..............................................................................................203
TRIGger:A:WIDTh:WIDTh...............................................................................................203
TRIGger:A:WIDTh:DELTa............................................................................................... 204
TRIGger:A:WIDTh:POLarity <Polarity>
Sets the polarity of the pulse trigger function.
Parameters:
<Polarity> POSitive | NEGative
POSitive
Positive going pulse, the width is defined from the rising to the
falling slopes.
NEGative
Negative going pulse, the width is defined from the falling to the
rising slopes.
*RST: POSitive
TRIGger:A:WIDTh:RANGe <RangeMode>
Defines how the measured pulse width is compared with the given limit(s) .
Parameters:
<RangeMode> WITHin | OUTSide | SHORter | LONGer
WITHin | OUTSide
Triggers on pulses inside or outside a range defined by time ±
delta. The time is specified with
TRIGger:A:WIDTh:WIDTh,
the range around is defined with
TRIGger:A:WIDTh:DELTa.
SHORter | LONGer
Triggers on pulses shorter or longer than a time set with
TRIGger:A:WIDTh:WIDTh.
*RST: LONGer
TRIGger:A:WIDTh:WIDTh <Time1>
For the ranges WITHin and OUTSide (defined using
TRIGger:A:WIDTh:RANGe), the
<Time1> defines the center of a range which is defined by the limits ±<Delta> (set with
TRIGger:A:WIDTh:DELTa).
For the ranges SHORter and LONGer, the width defines the maximum and minimum
pulse width, respectively.
Parameters:
<Time1> Center value, maximum value or minimum value depending on
the defined range type.
Range: 1.6E-08 to 3.435974E+01