Network Connections and Remote Operation
R&S
®
RTC1000
162User Manual 1335.7352.02 ─ 04
12.3 USB Connection
In addition to a LAN connection, you can use the USB connector at the rear panel to
access the instrument via USB.
Connecting the instrument using USB
1. Connect the USB cable to the USB type B connector at the rear panel of the instru-
ment, and to the computer.
2. Press the [SETUP] key.
3. Press "INTERFACE".
4. Press "USB" to select USB connection.
5. Press "PARAMETER".
6. Select the USB mode:
USB TMC (Test & Measurement Class)
USB VCP (Virtual Com Port)
USB MTP (Media Transfer Protocol)