Remote Commands Reference
R&S
®
RTC1000
263User Manual 1335.7352.02 ─ 04
Example:
EXT:DATA:POIN MAX
EXT:DATA:POIN?
Suffix: 20000
EXTern:DATA:HEADer?
Returns information on the external input signal waveform.
Table 13-5: Header data
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in Samples 20000
4 Number of values per sample interval, usually 1. 1
Return values:
<Header> Comma-separated value list.
Example: -9.477E-008,9.477E-008,20000,1
Usage: Query only
EXTern:DATA:XINCrement? <Xincrement>
Returns the time difference between two adjacent samples of the indicated waveform.
The command is relevant for data conversion, if binary data format is defined (FORM
UINT,8|16|32).
Parameters:
<Xincrement> Time in s
Usage: Query only
EXTern:DATA:XORigin? <Xorigin>
Returns the time of the first sample of the indicated waveform. The command is rele-
vant for data conversion, if binary data format is defined (FORM UINT,8|16|32).
Parameters:
<Xorigin> Time in s
Usage: Query only
EXTern:DATA:YINCrement? <Yincrement>
Returns the voltage value per bit of the indicated waveform. The command is relevant
for data conversion, if binary data format is defined (FORM UINT,8|16|32).
Parameters:
<Yincrement> Voltage in V
External input