Measurements
R&S
®
RTC1000
85User Manual 1335.7352.02 ─ 04
Figure 7-1: Cursor measurement menu
MEASURE TYPE
Defines the measurement type. To learn about the individual types, see
Chapter 7.2.2,
"Measurement Types"
, on page 86.
AUTOM. SOURCE, SOURCE
If "AUTOM. SOURCE" is activated ("On"), the selected channel is used as source for
the measurement. If the "AUTOM. SOURCE" is deactivated ("Off"), set the channel
with "SOURCE".
SET TO TRACE
Places the cursors in their optimal position on the waveform and allows fast and typi-
cally optimal automatic positioning of the cursors. Mostly, only fine-tuning is required. If
"SET TO TRACE" does not provide the anticipated results due to complex signals, use
the key "SET TO TRACE" to position the cursors in a predefined starting position, and
to return distant cursors to the screen.
SET TO SCREEN
Sets the cursors to their default position. Use this function to return distant cursors to
the screen.
GLUE TO TRACE
If active, the cursors stay on the selected data point without changing the position in
the measurement signal if the scaling is modified (cursors are "glued" to the signal).
The measured value remains unmodified.
Cursor Measurements