Rohde & Schwarz RTC1002EDU oscilloscope

Rohde & Schwarz RTC1000 manual - Page 148

For RTC1002EDU.

PDF File Manual, 371 pages, Read Online | Download pdf file

RTC1002EDU photo
Loading ...
Loading ...
Loading ...
Serial Bus Analysis
R&S
®
RTC1000
148User Manual 1335.7352.02 ─ 04
11.7.1 LIN Bus Configuration
Before you configure the bus, set the correct logic level (threshold).
For analog channels, see
"THRESHOLD menu" on page 36.
For digital channels, see
"To set the threshold for logic states" on page 108. The
default value is 500 mV.
To select the LIN bus
1. Press the [BUS] key in the Vertical section.
2. In the short menu, select the bus: "B1" or "B2".
3. In the "BUS" menu, press the "BUS TYPE" softkey.
4. Use the [Universal] knob to select "LIN":
To configure the CAN bus
1. In the "BUS" menu, press "CONFIGURATION".
2. Configure the bus. The settings are described below.
Figure 11-19: LIN bus configuration
LIN Bus (Option R&S RTC-K3)