Remote Commands Reference
R&S
®
RTC1000
175User Manual 1335.7352.02 ─ 04
Parameters:
<AcquisitionMode> RTIMe | ETIMe
RTIMe (Real Time Mode)
At slow time base settings the sampled points of the input signal
are used to build the waveform, no waveform points are added.
With fast time base settings, the sample rate is higher than the
ADC sample rate. Waveform samples are added to the ADC
samples with sin(x)/x interpolation.
ETIMe (Equivalent time)
The waveform points are taken from several acquisitions of a
repetive signal at a different time in relation to the trigger point.
*RST: RTIM
ACQuire:INTerpolate <Interpolation>
Defines the interpolation mode.
Parameters:
<Interpolation> LINear | SINX | SMHD
LINear
Linear interpolation between two adjacent sample points.
SINX
Interpolation by means of a sin(x)/x curve.
SMHD
Sample & Hold causes a histogram-like interpolation.
*RST: SINX
ACQuire:AVERage:COUNt <AverageCount>
Defines the number of waveforms used to calculate the average waveform. The higher
the number, the better the noise is reduced.
Parameters:
<AverageCount>
Only numbers from the 2
n
progression are permitted (2, 4, 8, ...)
Range: 2 to 1024
*RST: 2
ACQuire:AVERage:COMPlete?
Returns the state of averaging.
Acquisition and Setup