Rohde & Schwarz RTC1002EDU oscilloscope

Rohde & Schwarz RTC1000 manual - Page 248

For RTC1002EDU.

PDF File Manual, 371 pages, Read Online | Download pdf file

RTC1002EDU photo
Loading ...
Loading ...
Loading ...
Remote Commands Reference
R&S
®
RTC1000
248User Manual 1335.7352.02 ─ 04
MASK:ACTion:WFMSave:DESTination <File>
Defines the path and filename for a waveform data that will be saved on mask viola-
tion. The file format is CSV, the filename is incremented automatically
You can change the storage location, file name and/or file format manually in the [File]
> "Waveforms" menu. Remote control uses the recent settings.
Parameters:
<File> String parameter
Example:
MASK:ACT:WFMS:DEST "/USB_FRONT/MASKS/VIOL"
On first violation, the waveform data is saved to VIOL.CSV, on
second violation to VIOL01.CSV, the third to VIOL02.CSV ...
MASK:RESet:COUNter
Sets the counters of passed and failed acquisitions to zero.
Usage: Event
13.10 Function Generator
GENerator:FUNCtion......................................................................................................248
GENerator:VOLTage.......................................................................................................248
GENerator:VOLTage:OFFSet.......................................................................................... 249
GENerator:FREQuency.................................................................................................. 249
GENerator:FUNCtion:RAMP:POLarity..............................................................................249
GENerator:FUNCtion:PULSe:DCYCle..............................................................................249
GENerator:OUTPut[:ENABle].......................................................................................... 249
GENerator:FUNCtion <Function>
Selects the generator function.
Parameters:
<Function> DC | SINusoid | SQUare | PULSe | TRIangle | RAMP
*RST: SIN
GENerator:VOLTage <Amplitude>
Defines the amplitude value (peak-peak value) of the selected generator function.
Parameters:
<Amplitude> Range: 6.0000E-02 to 6.00000E+00
*RST: 5.0000E-01
Default unit: Vpp
Function Generator