Documenting Results
R&S
®
RTC1000
102User Manual 1335.7352.02 ─ 04
3.051758E+02,-1.19854E+01
4.577637E+02,-1.56854E+01
Import of CSV files: If you import a CSV file as reference waveform from a USB flash
drive to the instrument, the import converts the data to TRF format. The instrument
reads the first and the last time value and calculates the total time of the waveform,
and it counts the number of values. Then all amplitude values are read one by one and
written with an equidistant time distribution to the TRF file. If the first two time values
are identical, the waveform is considered to consist of pairs of values.
8.3.1.2 TXT Format
TXT files are ASCII files that contain only amplitude values but no time values. Ampli-
tude values are separated by commas. Pairs of values are listed as two subsequent
single values, without any identification. Amplitude values are given in scientific nota-
tion. There is no comma at the end of the file.
Amplitude values are given in scientific notation.
Example: TXT file
1.000E-02,1.000E-02,1.000E-02,1.000E-02,3.000E-02
8.3.1.3 BIN Format
BIN files contain only binary amplitude values but no time values. Each value has a
word size of 8, or 16, or 32 bit, the word size is the same throughout the file.
You can set the word order: BIN MSBF saves data in Big Endian order - beginning with
the MSB (Most Significant Byte) and ending with the LSB (Least Significant Byte). BIN
LSBF saves data beginning with the LSB and ending with the MSB. Pairs of values are
listed as two subsequent single values, without any identification.
8.3.1.4 FLT Format
FLT files contain amplitude values in float format, where 4 successive bytes are saved
in a 32-bit float value.
You can set the word order: FLT MSBF saves data in Big Endian order - beginning with
the MSB (Most Significant Byte) and ending with the LSB (Least Significant Byte). FLT
LSBF saves data beginning with the LSB and ending with the MSB.
8.3.1.5 TRF Format
TRF is the specific binary format for reference waveforms of the R&S RTC1000. It con-
tains the amplitude value of each sample that is displayed on the screen (8 bit or 16 bit
long). For peak-detect waveforms, 2 values per sample are saved. The file contains
also time information (time of the first sample and the sample interval) and current
instrument settings.