Remote Commands Reference
R&S
®
RTC1000
195User Manual 1335.7352.02 ─ 04
POD:DATA:HEADer?
Returns information on the digital channel waveform.
Position Meaning Example
1 XStart in s -9.477E-008 = - 94,77 ns
2 XStop in s 9.477E-008 = 94,77 ns
3 Record length of the waveform in samples 200000
4 Number of values per sample interval, usually 1. 1
Return values:
<Header> Comma-separated value list.
Example:
-9.477E-008,9.477E-008,200000,1
Usage: Query only
POD:DATA:POINts <PointSelection>
As a setting, the command selects a range of samples that will be returned with
POD:
DATA?
. As a query, the command returns the number of returned samples for the
selected range. If
ACQuire:WRATe is set to MSAMples (maximum sample rate), the
memory usually contains more data samples than the screen can display. In this case,
you can decide which data will be saved: samples stored in the memory or only the
displayed samples.
Parameters:
<PointSelection> DEFault | MAXimum | DMAXimum
Sets the range for data queries.
DEFault
Waveform samples displayed on the screen.
MAXimum
Range is the complete memory
Note: Reading the complete memory is only possible in STOP
mode, and if maximum sample rate is set (
ACQuire:WRATe
ACQ:WRAT MSAM
)
DMAXimum (Display Maximum)
Number of samples in the current waveform record.
*RST: DEFault
Return values:
<Points> Number of data points in the selected range.
Default unit: Samples
See also:
CHANnel<m>:DATA:POINts
Acquisition and Setup
Loading ...
Loading ...
Loading ...