continuously or enable the touch screen to select it. You can select CH1-CH4 or
D0-D15.
- When you select the analog channel (CH1-CH4), press Threshold, then rotate
the multifunction knob or use the pop-up numeric keypad to set the
threshold of the source channel. When you modify the threshold of the clock
channel, a dotted line displaying the current threshold level is displayed on the
screen. It disappears in about 2 s after you stop modifying the threshold.
4. Set the Signal
Press Settings to enter the signal setting menu.
- Select the channel
Press Channel to select a channel ("A" or "B") that matches the actual FlexRay
bus signal.
- Specify the signal rate
Press Baud to open the baud rate list (2.5 Mbps, 5 Mbps, and 10 Mbps), and
select a preset baud rate that matches the actual FlexRay bus signal. You can
also press Baud continuously, then rotate the multifunction knob or use the
numeric keypad to set the desired baud rate.
- Set the signal type
Press Signal and rotate the multifunction knob to select a signal type that
matches the actual FlexRay bus signal. Press down the knob to select it. You can
also press Signal continuously or enable the touch screen to select it. The
available signal types include BP, BM, and RX/TX.
- Sample position
Sample position is a point within a bit’s time. The oscilloscope samples the bit
level at this point. The sample position is represented by the proportion of "the
time from the start of the bit to the sample point" to the "bit time", as shown in
the figure below.
Protocol Decoding