- Press Source, rotate the multifunction knob to select the desired source
channel, and then press down the knob it. You can also press Source
continuously or enable the touch screen to select it. You can select CH1-CH4 or
D0-D15.
- When you select the analog channel (CH1-CH4), press Threshold, then rotate
the multifunction knob or use the pop-up numeric keypad to set the
threshold of the source channel. When you modify the threshold of the clock
channel, a dotted line displaying the current threshold level is displayed on the
screen. It disappears in about 2 s after you stop modifying the threshold.
4. Set the Signal
Press Settings to enter the signal setting menu.
- Set the baud rate
Three methods are available for setting the baud rate:
- Press Baud Rate, then set the user-defined baud rate with the displayed
numeric keypad.
- Press Baud Rate to select the preset baud rate. The available baud rates
include 2.4 kbps, 4.8 kbps, 9.6 kbps, 19.2 kbps, and etc.
- Press Baud Rate, then rotate the multifunction knob to baud rate at a
small step of 1 bps.
- Set the parity bit
Press Parity bit continuously to select whether the data contain the parity bit.
- Set the protocol version
Press Version, and rotate the multifunction knob to select a protocol version
that matches the LIN bus signal. Then, press down the knob to select it. You can
also press Version continuously or enable the touch screen to select it. The
available versions include 1.X, 2.X, and Both.
5. Set Display-related Parameters
Press Display to enter the display setting menu.
- Set the display format
Protocol Decoding