T
Source A
Source B
Edge A
Edge B
Figure 8.10 Delay Trigger
Trigger Type:
Press Type, and then rotate the multifunction knob to select "Delay". Press down
the knob to select the trigger type. Then, the current trigger setting information is
displayed at the upper-right corner of the screen, as shown in the figure below. You
can also press Type continuously or enable the touch screen to select it.
Source A:
Press SourceA to select CH1-CH4 or D0-D15 as the trigger source of Source A. For
details, refer to descriptions in "
Trigger Source
". The current trigger source is
displayed at the upper-right corner of the screen.
NOTE
Only when we select the channel (that has been input with signals) as the trigger source, can
we obtain a stable trigger.
Edge A:
Press
EdgeA continuously to select the trigger edge type of Source A in Delay trigger.
It can be set to "Rising" or "Falling" .
Source B:
Press SourceB to select CH1-CH4 or D0-D15 as the trigger source of Source B. For
details, refer to descriptions in "
Trigger Source
". The current trigger source is
displayed at the upper-right corner of the screen.
NOTE
Only when we select the channel (that has been input with signals) as the trigger source, can
we obtain a stable trigger.
Edge B:
To Trigger the Oscilloscope
104
Copyright ©RIGOL TECHNOLOGIES CO., LTD. All rights reserved.
Loading ...
Loading ...
Loading ...