Rigol MSO8204A oscilloscope

Rigol MSO8000A series user guide - Page 103

For MSO8204A.

PDF File Manual, 406 pages, Read Online | Download pdf file

MSO8204A photo
Loading ...
Loading ...
Loading ...
• When you select "Rising" for the edge type and "<" for trigger condition, the
oscilloscope triggers when the positive slope time of the input signal is smaller
than the specified time.
• When you select "Rising" for the edge type and "<>" for trigger condition, the
oscilloscope triggers when the positive slope time of the input signal is greater
than the specified lower limit time and smaller than the specified upper limit
time.
• When you select "Falling" for the edge type and ">" for trigger condition, the
oscilloscope triggers when the negative slope time of the input signal is greater
than the specified time.
• When you select "Falling" for the edge type and "<" for trigger condition, the
oscilloscope triggers when the negative slope time of the input signal is smaller
than the specified time.
• When you select "Falling" for the edge type and "<>" for trigger condition, the
oscilloscope triggers when the negative slope time of the input signal is greater
than the specified lower limit time and smaller than the specified upper limit
time.
Slope Time Setting:
• When the Trigger Condition is set to ">" or "<", press Lower or Upper, and
then rotate the multifunction knob to set the lower limit value or the upper
limit value. You can also use the numeric keypad to set it. The available range of
the slope time is from 800 ps to 10 s.
•
When the
Trigger Condition is set to "<>", press Upper and Lower, and then
rotate the multifunction knob to set the upper limit value and the lower limit
value. You can also use the numeric keypad to set them.
NOTE
The lower slope time limit must be smaller than the upper slope time limit.
Level Selection and Adjustment of Trigger Level:
To Trigger the Oscilloscope