NOTE
The default values for threshold upper limit, threshold middle value, and threshold low limit
are 90%, 50%, and 10%, respectively. You can press > Setting to enter the sub-menu to
set the parameters.
Count Values
1. Positive Pulse Count: The number of positive pulses that rise from under the
threshold lower limit to above the threshold upper limit.
Positive Pulse Count = n
1 n2
Threshold
Upper Limit
Threshold
Lower Limit
2. Negative Pulse Count: The number of negative pulses that fall from above the
threshold upper limit to below the threshold lower limit.
1 n2
Negative Pulse Count = n
Threshold
Upper Limit
Threshold
Lower Limit
3. Rising Edge Count: The number of rising edges that rise from under the threshold
lower limit to above the threshold upper limit.
Operations and Measurements