Siglent SDS6000L Low Profile Digital Oscilloscope

User Manual - Page 95

For SDS6000L.

PDF File Manual, 310 pages, Read Online | Download pdf file

SDS6000L photo
Loading ...
Loading ...
Loading ...
SDS6000L User Manual
int. s i g l e n t . c o m 93
Less than a time value
(
≤
)
-- Trigger when the positive or negative pulse time of the input signal is
lower than the specified time value. Below is an example of a trigger condition set to a positive pulse
width < 100ns
Great than a time value
(≥)-- Trigger when the positive or negative pulse time of the input signal is
greater than the specified time value. Below is an example of a trigger condition set to a positive pulse
width > 100ns
Within a range of time values
(
[--,--]
)
-- Trigger when the positive or negative pulse time of the input
signal is greater than the specified lower limit of time and lower than the specified upper limit of the
time value. Below is an example of a trigger condition set to 100 ns < positive pulse width < 300 ns.
Outside a range of time value
(
--][--
)
-- trigger when the positive or negative pulse time of the input
signal is greater than the specified upper limit of time and lower than the specified lower limit of the
time value.
Holdoff, coupling, and noise reject can be set in pulse trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.