2 2 6 i n t . s i g l e n t . c o m
25 Histogram
25.1 Overview
The SDS6000L supports waveform histograms for observing probability distributions of the waveform
in the specified region. The statistics can be performed in both horizontal and vertical directions. The
histogram continues to update as long as the acquisition is active.
Click Analysis > Histogram to recall the Histogram dialog box:
A. Turn on or off the histogram
B. Select the source of the histogram:
C1~C8. When Zoom is turned on, the
source selection is automatically
switched to Z1-Z8
C. Set the type of histogram: Horizontal,
Vertical or Both
D. Set the region of the histogram. The
setting range is within the waveform
area.
E. Turn on or off histogram statistics
F. Clear and restart the histogram
statistics.
Type
 Horizontal – Displays the histogram in the horizontal direction. The oscilloscope counts waveform
data falling into every horizontal (time) bin defined by the histogram region and shows the
histogram at the bottom of the waveform area.
 Vertical – Displays histogram in the vertical direction. The oscilloscope counts waveform data
falling into every vertical (amplitude) bin defined by the histogram region and shows the histogram
at the left of the waveform area.
 Both – Displays both horizontal and vertical histograms