Siglent SDS6000L Low Profile Digital Oscilloscope

User Manual - Page 227

For SDS6000L.

PDF File Manual, 310 pages, Read Online | Download pdf file

SDS6000L photo
Loading ...
Loading ...
Loading ...
SDS6000L User Manual
int. s i g l e n t . c o m 2 2 5
24.2 Mode
The counter provides 3 modes. Click mode in the counter dialog box to open the mode selection
window:
ï‚Ÿ
Frequency:
Average frequency over a set period
ï‚Ÿ
Period:
The reciprocal of the average frequency over a set period
ï‚Ÿ
Totalizer:
Cumulative count
When the mode is set to totalizer, the edge of the counting source needs to be set. When the counting
gate is turned on, the counter can count only when the gate-source meets the specified conditions.
Click Gate Setting to recall the gate setting dialog box:
A. Turn on or off the gate
B. Select the gate type: Level or After Edge
C. Gate source display area. C1 and C2 are gate sources of
each other, C3 and C4 are gate sources of each other, C5
and C6 are gate sources of each other, C7 and C8 are gate
sources of each other
D. When the gate type is level, set the polarity (Positive or
Negative) of the gate source. When the gate type is edge,
set the slope (Rising or Falling edge) of the gate source.
E. Set the gate level
F. Return to the previous menu