SDS6000L User Manual
2 8 2 i n t . s i g l e n t . c o m
Zero Adjust
The output zero error of the AWG can be calibrated by using auto mode or manual mode. The object
of the Zero Adjust is: when AWG outputs a 0 V DC signal, the measured mean value of the
corresponding channel is within ± 1mV.
Auto Steps
: In this mode, the adjustment can only be performed with C1 of the oscilloscope.
Connect the output of AWG to C1 of the oscilloscope, click Auto , and the AWG will start to
automatically adjust. The scope will display Zero adjust completed!” when the adjustment is done.
Manual Steps
: The AWG can be manually calibrated through any channel on the oscilloscope.
Take C2 as an example:
1. Connect the output of AWG to C2, open C2, set it to DC coupling, turn on the bandwidth limit,
and set the probe attenuation to 1X.
2. Set the vertical scale of C2 to a small scale such as 1 mV/div. Turn on the measure and set
the parameter to the Mean of C2.
3. Click Manual and roll the mouse wheel to adjust the compensation value, until the mean
value of C2 is within ± 1 mV, and then click Save .
31.4 System