SDS6000L User Manual
11 8 i n t . s i g l e n t . c o m
Copy Setting
Click the Copy Setting in the decode dialog
box to synchronize the settings between the
trigger and decode.
A. Copy the decode settings to trigger
B. Copy the trigger settings to decode
C. Return to the previous menu.
Note:
The synchronization is not automatic. If the settings in one place
change, a copy operation is necessary to re-synchronize.
14.2.2 I2C Trigger
When the protocol is set to I2C, the following trigger conditions can be set: Start, Stop, Restart, No Ack,
EEPROM, or an R/W frame with a specific device address and data value.
Click Trigger Setting in the
I2C trigger dialog box to select
the trigger condition:
Start
-- The oscilloscope will be triggered when the SDA line transitions from high to low while the SCL
is high.