Siglent SDS6000L Low Profile Digital Oscilloscope

User Manual - Page 69

For SDS6000L.

PDF File Manual, 310 pages, Read Online | Download pdf file

SDS6000L photo
Loading ...
Loading ...
Loading ...
SDS6000L User Manual
int. s i g l e n t . c o m 67
11.2 Acquisition Setup
11.2.1 Overview
Click Acquire Menu on the quick menu of the timebase settings, or click the menu bar
Acquire > Menu to recall the Acquire dialog box on the right side.
A. Select the interpolation mode
B. Select the Acq mode
C. Select the acquisition mode (Normal / Peak /
Average / ERES). See the section "Acquisition"
for details.
D. Select the Memory Management mode (Auto,
Fixed Sample Rate, and Fixed Memory). See
the section "Memory Management" for details.
E. Select the maximum memory depth
F. Enter the sequence menu. See the section
Sequencefor details
G. ESR mode on/off. See the section “ESR for
details
H. XY mode on/off
Interpolation
-- At small timebase settings, the number of original points on the screen may be less
than the number of display pixels in the grid area, so interpolation is necessary to display a continuous
waveform. For example, at 1 ns/div timebase and 5 GSa/s sample rate, the number of original points
is 50, but the grid area includes 1250 horizontal pixels. In this case, the oscilloscope needs to
interpolate the original points by 25.