If you set the trigger condition to 7 address & data or 10
address & data:
ï‚Ÿ Address can be selected in the hexadecimal
range of 0x00 to 0x7F (7-bit) or 0x3FF (10-bit). If the
address is selected as "0xXX (7-bit address)" or
"0xXXX (10-bit address)", the address is ignored. It
will always trigger on the Ack bit followed by the
address.
ï‚Ÿ Data1 and Data2 can be selected in the
hexadecimal range of 0x00 to 0xFF. If the data is
selected as "0xXX", the data is ignored. It will
always trigger on the Ack bit followed by the
address.
ï‚Ÿ R/W Bit can be specified as Write, Read, or Don't
Care.
Data Length
-- The data length range is 1 to 12 bytes. When the SDA data length is equal to the value
of Data Length and address length is the same as the set value, the oscilloscope will be triggered.
ï‚Ÿ Click the Address Length to select "7-bits" or "10-bits" to match the address of the input signal.
ï‚Ÿ Click the Data Length and roll the mouse wheel or use the virtual keypad to set the data length
to match the data length of the input signal.