SDS6000L User Manual
int. s i g l e n t . c o m 2 1 3
Perform Analysis > Mask Test to open the Mask Test dialog box:
A. Turn on/off the test
B. Select the source
C1~C8
C. Select the rule
All In, All Out, Any In, and Any Out
D. Set the mask
E. Change the state of the Pass/Fail operation. Turning
off the operation when the test is in progress will stop
the test immediately, and the counters in the
information display area will stop. Turning the test
back on will re-start the test and clear the counters.
F. Turn on/off the information display
G. Turn on/off Stop on Fail. When it is "on", the
oscilloscope stops the acquisition once it detects a
failure
H. Turn on/off “Failure to History”. When it is "on", the
failure frame will be stored internally and can be
viewed through the history function
I. Turn on/off “Capture on Fail”. When it is "on", the
failure waveform is detected, and the screenshot with
the failure frame will be stored in the external U disk.
J. Turn on/off the sound prompt when a failure occurs.