Set the values for Mask X and Mask Y (in divisions of the display graticule), and then perform
Create Mask to generate the mask. The horizontal and vertical adjustment range is 0.08~4.00 div.
X = 0.2 div, Y = 0.2 div
X = 1 div, Y = 1 div
Saving and recalling mask files (*.msk) is similar to the operation of setup files, see the chapter
"Save/Recall" for details.
22.2.2 Mask Editor
The Mask Editor is a built-in tool that provides an approach to create custom masks. Below is its layout:
A. Menu bar
B. Coordinate the latest selected point on the display
C. Mask edit area, which is equivalent to the grid area. In this example, a hexagon has been