Trigger source, slope (rising, falling), limit range, and time value can be set in the trigger dialog box.
Holdoff, coupling, and noise reject can be set in interval trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.
13.5.8 Dropout Trigger
Dropout trigger includes two types: Edge and State.
Edge
Trigger when the time interval (
â–³
T) from when the rising edge (or falling edge) of the input signal
passes through the trigger level to when the neighboring rising edge (or falling edge) passes through
the trigger level is greater than the set time, as shown in the figure below:
State
Trigger when the time interval (
â–³
T) from when the rising edge (or falling edge) of the input signal
passes through the trigger level to when the neighboring falling edge (or rising edge) passes through
the trigger level is greater than the set time, as shown in the figure below:
Trigger source, slope (rising, falling), dropout type, and time value can be set in the trigger dialog box.