int. s i g l e n t . c o m 1 0 3
The logical setting of analog channels
The logical setting of digital channels
Limit Range
This setting is particularly useful to filter the hazard signals of combinational logic.
Holdoff can be set in pattern trigger, see the section "Holdoff" for details.
13.5.11 Qualified Trigger
The qualified trigger is an edge trigger after a user-defined qualified condition is satisfied. So a qualified
trigger has two sources; one is the source of the edge trigger, and the other is the qualified source.
The qualified types include “State”, “State with Delay”, “Edge”, and “Edge with Delay”. When the type
is “State”, the oscilloscope triggers on the first edge when the qualified source is in the specified state
(High or Low). When the type is “State with Delay”, a time limit condition is also available.
Qualified
Source
Edge
Trigger
Source
Trigger
Position
Qualified State = High
Edge = Rising