SDS6000L User Manual
int. s i g l e n t . c o m 99
To set window trigger via the Relative window type
When the window trigger type is set to "Relative",
click the trigger descriptor box. The pop-up
menu will show two user-defined parameters:
Level +/-Delta" and "Level Center".
The above two parameters can be set in the
following way:
Select the parameter in the Level +/-Delta area of the quick menu, then set the parameter value by
the virtual keypad or the mouse wheel. Setting the center level is similar.
Note:
"Level +/-Delta” represents half of the actual window area. For example,
when the value is 200mV, it represents a range of ±200 mV, which is a 400mV
window.
Holdoff, coupling, and noise reject can be set in the window trigger, see the sections "Holdoff", "Trigger
Coupling" and "Noise Reject" for details.
13.5.7 Interval Trigger
Trigger when the time difference between the neighboring rising or falling edges meets the time limit
condition.
When the trigger condition is set as an interval between two neighboring rising edges and it is less than
the set time value, the trigger diagram is as follows: