SDS6000L User Manual
2 9 8 i n t . s i g l e n t . c o m
A. Check Automatic to enable dynamic IP. In
this case, the oscilloscope must be connected
to the local area network with a DHCP server.
Contact your network administrator to confirm
the relevant information.
B. When is not checked, the oscilloscope
uses static IP. Set the static IP address,
subnet mask, and gateway separately.
C. If your network provides Dynamic DNS, you
can enable Automatic (DNS) to let the
oscilloscope register its hostname and use the
DNS server for name resolution.
D. When is not checked, set the DNS IP
address manually.
E. When accessing more than two SIGLENT
instruments through the web browser, it is
necessary to set a different VNC port number
for each instrument. The range is from 5900 to
5999.
33.3.2 Clock Source
The clock source of the device can be set to "Internal" or "External". When the clock source is set to
"external", the device receives the 10 MHz clock from the "10 MHz In" port on the rear panel and uses
it as a reference to generate the sampling clock of the system. Both “internal" and "external" clock
sources can be output from the 10 MHz Out port for synchronizing other instruments.
When the clock source is set to "External", an icon indicating an external clock will appear in the lower
right corner of the display:
The clock source is external.
The clock source is external, but no valid external clock is detected.