14.9.2 I2S Trigger
Click Trigger Setting in the I2S trigger dialog box to set the trigger conditions :
A. Audio Variant: Audio-I2S, Audio-LJ,
Audio-RJ
B. Trigger Condition: Data, Mute, Clip,
Glitch, Rising Edge, Falling Edge
C. Channel: Set the channel to trigger, Left
or Right
D. Set the bit order to MSB or LSB
E. Channel Bits: Specify the receiver word
length. The range is from 1 to 32 bits
F. Start Bit: The start bit of data, ranging
from 0 to 31
G. Data Bits: Specify the number of bits per
word. It should be less than the channel
bits
H. When the "Trigger Condition" is Data, set
the compare type to =, >, <
I. When the "Trigger Condition" is Data, set
the data value
J. Return to the previous menu
Trigger Condition
ï‚Ÿ
Data
-- Trigger on data.
✓ Click Compare Type to select "=", ">" or "<".
✓ Click Value to set the data value by the mouse wheel or virtual keypad. The range of data
value is related to the number of Data Bits .
ï‚Ÿ
Mute
--
Trigger on the mute signal. Mute signal: Volume is less than the set value and duration
reaches the set value.