C. Set the sweep frequency. The frequency mode can be linear or decade. When it is set to
linear, the corresponding center frequency and span frequency need to be set; when it is set
to a decade, the corresponding start frequency and stop frequency need to be set
D. Set the number of sweep points. The larger the number of points, the higher the sweep
resolution
E. Set the amplitude of the sweep signal
F. Set the offset of the sweep signal
G. Set the signal amplitude unit. When it is set to dB, the reference level and load need to be
set
H. Set the load
A. Set the sweep type to Vari-Level
B. Select a profile. Up to 4 profiles can be edited
C. Set the frequency mode
D. Set the number of sweep points
E. Set the offset of the sweep signal
F. Set the signal amplitude unit. When it is set to dB, the reference level and load need to be
set
G. Variable level signal display area
H. Click Edit to turn on the profile editor