int. s i g l e n t . c o m 63
The quick menu only covers the height range and position of the digital channels display area. More
settings can be found in the dialog box.
A. Turn on/off the digital channels
B. Labels can be set to data, address, or custom
characters.
C. Logic threshold of D7~D0. The oscilloscope will
automatically set the threshold according to the
specified logic family, or the user can set the
threshold manually using the Custom option.
D. Logic threshold of D15~D8
E. Set the channels to be displayed
F. Deskew setting, the same as setting analog
channels
G. Bus setting
Logic Threshold Setting
The threshold level determines how the input signal is evaluated. The threshold level can be set in the
Logic Setting . The input voltage less than the threshold is recognized as a '0', and the input voltage
greater than the threshold is recognized as a ‘1’.