DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
89
displayed on the rear panel [Aux Port] will disappear immediately.
Off: When the oscilloscope detects a waveform that matches the set “Mode”, the test
continues. The information displayed on the screen continues to update in real time. If
“On” is selected for “Output”, the [Aux Port] on the rear panel will output a pulse.
Press the Output soft key to select whether to output the test results. When the oscillo-
scope detects a waveform that matches the set “mode”, the Aux port on the rear panel
outputs a waveform; when it does not detect a waveform that matches the set “mode”,
there is no output.
Press the Mode soft key:
Pass: When a pass is detected, the statistical results and output pulses are displayed,
and the buzzer does not alarm.
Failure: When a failure is detected, the statistical results and output pulses are dis-
played, and the buzzer does not alarm.
Pass ring: When one pass is detected, the statistical result and output pulse are dis-
played, and the buzzer alarms (not related to the sound switch state).
Failure ring: When a failure is detected, the statistical result and output pulse are dis-
played, and the buzzer alarms (not related to the sound switch state).
Press the Output to DDS soft key to select the test result output path.
Select the output to the signal source (GEN1 / GEN2 / GEN3, the oscilloscope auto-
matically turns on the burst enable of the corresponding signal source, and set the
"Source" to "Pass / Fail", the user only needs to manually enable the corresponding
signal source to enable) or directly Output from the AUX port on the rear panel (off, the
user need to manually set "Utility"-> "AUX" to select Pass / Fail).
For the setting of the signal source, please refer to section Burst.
Segmented acquisition
Segmented storage when multiple triggers are made in the acquisition process, the da-
ta obtained for each trigger sampling is stored in the storage space of each segment.
Segment storage principle
As shown in the figure below, the data collected when the first trigger occurs is stored
in the first segment of storage space, the data collected when the second trigger occurs
is stored in the second segment of storage space, and so on, until When the number of
segments set by the user is acquired, the oscilloscope enters the pause state to end
the acquisition process. Read the first segment of the acquisition for display.