Loading ...
Loading ...
Loading ...
![](https://files.manualsfile.com/1694261-dpo6004bc/bg73.png)
DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
112
Dynamic range
± 5.0V + threshold
Minimum voltage
swing
500 mVpp
Vertical resolution
1 bit
Trigger
Trigger level range
± 4 divisions from the center of the screen
Trigger mode
Auto, Normal, Single
Holdoff range
8ns~10s
Trigger level accu-
racy
CH1~CH4 0.2 divisions × volts / division within ± 4 divisions from the
center display
Edge trigger
Slope
Rising, Falling, Rising or Falling
Data source
CH1~CH4,
D1.0~D1.3,
D2.0~D2.3,
D3.0~D3.3,
D4.0~D4.3
Pulse width trigger
Polarity
Positive, Negative
Conditions
<, >, !=, =
Data source
CH1~CH4,
D1.0~D1.3,
D2.0~D2.3,
D3.0~D3.3,
D4.0~D4.3
Pulse width range
8ns~10s
Video trigger
Signal standard
NTSC, PAL, HDTV720P, HDTV1080P, HDTV1080i