DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
109
The oscilloscope must fall within the factory calibration period.
All specifications are guaranteed to meet the requirements except those marked as
"Typical."
Oscilloscope technical specifications
Acquire
Acquisition meth-
od
Normal, average, peak, high resolution
Real-time
sampling rate
1 G Sa/s (single channel)
500 M Sa/s (dual channel)
250 M Sa/s (three / four channels);
Peak Detect
Analog Channel 4ns
Averaging mode
N samples for analog channels (N values: 2, 4, 8, 16, 32, 64, 128, 256,
512, and 1024)
High resolution
12bit
Minimum detection
pulse width
8ns
Memory Depth
Single Channel 64M / 128M [EDU]
Dual channel 32M / 64M [EDU]
Three or four channels 16M / 32M [EDU]
Input
Number of chan-
nels
4 analog channels, 16 (4 * 4) digital channels
The number of analog and digital channels opened at the same time
should be less than 4.
Input coupling
DC, AC or GND
Digital channels are not supported
input resistance
DC coupling
Analog channel: 25pF ± 3pF, 1MΩ ± 2%
Digital channels: (300KΩ ± 2%), (8 pF ± 3 pF)
Support probe at-
tenuation coeffi-
cient
Analog channels 0.01X, 0.02X, 0.05X, 0.1X, 0.2X, 0.5X, 1X, 2X, 5X, 10X,
20X, 50X, 100X, 200X, 500X, 1000X, 2000X, 5000X, 10000X
Voltage rating
300V CAT II