Hantek DPO6004BC series oscilloscope

Hantek DPO6004B user manual - Page 42

For DPO6004BC.

PDF File Manual, 124 pages, Read Online | Download pdf file

DPO6004BC photo
Loading ...
Loading ...
Loading ...
DPO6000, MPO6000 Series Digital Fluorescent Oscilloscope Product Manual V1.3
39
trigger slope to rising edge.
Pulse width trigger
Pulse width trigger sets the oscilloscope to trigger on a positive or negative pulse of a
specified width. You can set the trigger source, polarity (positive pulse width, negative
pulse width), limit conditions, and pulse width in this menu.
Trigger point
Trigger level
Positive pulse width Negative pulse width
Trigger point
Press the front panel Trig Menu button to enter the trigger function menu.
[Type] Select the pulse width and press V0 to confirm.
[Data source] Select CH1CH4 or LA as the trigger source.
Note: LA must be inserted when LA is used as the trigger source.
[Polarity] Select the positive or negative polarity of the trigger.
[When] Select the trigger conditions [, , , ], And press V0 to confirm.
[Greater than the set time value]: When the positive or negative pulse width of the
input signal is greater than the set pulse width, it can be triggered (pulse width error is
5%).
[Smaller than the set time value]: When the positive or negative pulse width of the
input signal is smaller than the set pulse width, it can be triggered (pulse width error is
5%).
[Equal to the set time value]: When the positive or negative pulse width of the input
signal is equal to the set pulse width, it can be triggered (pulse width error is 5%).
[Not equal to the set time value]: When the positive or negative pulse width of the
input signal is not equal to the set pulse width, it can be triggered (pulse width error is
5%).
[Width] Set the pulse width time from 8ns to 10s.
[50%] Set the trigger level to the vertical midpoint of the peak-to-peak value of the trig-
ger signal. The trigger level value is displayed in the upper right corner of the screen.
[Mode] Select the acquisition mode [Auto, Normal] and press V0 to confirm.
[Holdoff] Set the holdoff time.
Loading ...
Loading ...
Loading ...