SDS1000X-E User Manual
68
WWW.SIGLENT.COM
Figure 29 Slope Trigger
6. Press the
Limit Range
softkey; then turn the
Universal Knob
to select the desired
slope condition, and push down the knob to confirm.
<= (less than a time value):
trigger when the positive or negative slope time of
the input signal is lower than the specified time value.
>= (greater than a time value):
trigger when the positive or negative slope time of
the input signal is greater than the specified time value.
[--,--] (within a range of time value):
trigger when the positive or negative slope
time of the input signal is greater than the specified lower limit of time and lower
than the specified upper limit of time value.
--][-- (outside a range of time value):
trigger when the positive or negative slope
time of the input signal is greater than the specified upper limit of time and lower
than the specified lower limit of time value.
Coupling and noise reject can be set in slope trigger, see the sections "Trigger Coupling"
and "Noise Rejection" for details.