User Manual - Page 196

For SDS1202X-E.

Loading ...
Loading ...
Loading ...
SDS1000X-E User Manual
171
WWW.SIGLENT.COM
7. Press the
X Mask
or
Y Mask
softkey; and then turn the
Universal Knob to
select the
desired value. The range is from 0.02div to 4div.
8. Press the
Create Mask
button to create the mask. Whenever the
Create Mask
softkey
is pressed the old mask is erased and a new mask is created.
9. Press the
UP
softkey to return to the PASS/FAIL function menu.
10. Press the
Msg Display
to select
On
or
Off
to turn on or off the message display. When
On
is selected, the test result will be displayed in the upper-right message box of the
screen.
F:
the frame number of the channel waveform that fails to match the mask.
P:
the frame number of the channel waveform that matches the mask.
T:
the frame number of the channel waveform that have been acquired. It is the
total of
Pass
and
Fai
l.
11. Press
the Next Page
softkey to go to the second page of the PASS/FAIL function
menu.
12. Press the
Stop On Fail
softkey to select
On
or
Off
to turn on or off the function.
On:
when failed waveforms are detected, the oscilloscope will stop the test and
enter the STOP state. At this point, the results of the test remain the same on the
screen (if display is turned on) and only one pulse is output from the
[TRIG OUT]
connector (if enabled) at the rear panel.
Off:
the oscilloscope will continue with the test even though failed waveforms are
detected. The test results on the screen will update continuously and a pulse will
be output from the
[Trigger Out]
connector at the rear panel each time a failed
waveform is detected.
13. Press the
Output
softkey to select turn on or off the sound.
: When the failed waveforms are detected, the beeper sounds.
: The beeper will not sound even if the failed waveforms are detected,
14. Press the Next Page softkey to return to the first page of the PASS/FAIL function
menu.
15. Press the
Operate
softkey to perform the test.
: Current state is stop; press the softkey to start the pass/fail test.
: Current state is running; press the softkey to stop the pass/fail test.
Save and Recall Test Mask
Users can save the current test mask to the internal Flash memory or external USB flash
device. The file format of the test mask file is “*.msk”.
Save Test Mask to Internal Memory
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
Loading ...
Loading ...
Loading ...