User Manual - Page 195

For SDS1202X-E.

Loading ...
Loading ...
Loading ...
SDS1000X-E User Manual
170
WWW.SIGLENT.COM
Pass/Fail Test
One way to verify a waveform's compliance to a particular set of parameters is to use
pass/fail testing. A pass/fail defines a region of the oscilloscope's display in which the
waveform must remain in order to comply with chosen parameters. Compliance to the
mask is verified point- by- point across the display.
The test results can be displayed on the screen as well as be declared through the system
sound or the pulse signal output from the
[TRIG OUT]
connector at the rear panel.
Pass/Fail test operates on displayed analog channels; it does not operate on channels
that are not displayed.
Figure 110 Pass/Fail Test
Set and Perform Pass/Fail Test
Do the following steps to set and perform pass/fail test:
1. Press the
Utility
button on the front panel to enter the UTILITY function menu.
2. Press the
Next Page
softkey to go to the second page of the UTILITY function menu.
3. Press the
Pass/Fail
softkey to enter the PASS/FAIL function menu.
4. Press the
Enable Test
softkey to select
On
to enable the pass/ fail test.
5. Press the
Source
softkey to select the desired channel.
6. Press the
Mask Setting
softkey to enter the MASK function menu.