User Manual - Page 218

For SDS1202X-E.

Loading ...
Loading ...
Loading ...
SDS1000X-E User Manual
193
WWW.SIGLENT.COM
Bode Plot II
The bode plot application for the SDS1000X-E (only 4 channels scope supports this
function) controls an external arbitrary waveform generator(only Siglents arbitrary
waveform generator can be supported) to sweep a sine wave across a range of
frequencies while measuring the input to and output from a device under test (DUT). At
each frequency point, gain and phase are measured and displayed on the Bode plot.
Perform Bode Plot
The following steps briefly describe how to perform the Bode Plot:
1. Press the
Utility
button on the front panel to enter the UTILITY function menu, and
then press
the Next Page
softkey to enter the second page.
2. Press
Bode Plot II
softkey to enter the BODE PLOT II menu. At this time, only the
Default
and
Auto Setup
button which on the front panel are available.
Figure 127 BODE PLOT II Menu
3. Configure the parameters (see the section Setting for details).
4. Press the
Run
softkey to perform the test.
Setting
1. Press the
Config
softkey to enter the CONFIGURE menu. All the parameters in this
menu are used to configure the external AWG.
Figure 128 CONFIGURE Menu
1) Press the
Source
softkey to enter SOURCE menu.
Press the
Interface
softkey and turn the
Universal Knob
to select the
connection mode of AWG.
USB Connect the oscilloscope USB host to the external AWG
s USB
Device via a USB cable.
Loading ...