SDS1000X-E User Manual
190
WWW.SIGLENT.COM
Reference Position
The reference position setting determines the physical point that the oscilloscope uses
during vertical and horizontal scale changes. In some situations, it is more convenient to
use a fixed position on the display.
Press the
Utility
button on the front panel and then press the
Reference Pos.
softkey to
enter the Reference POS menu
.
Press the
Vertical
softkey to select
Fixed Offset
or
Fixed Position
.
Fixed Position
: when the vertical gain is changed, the oscilloscope will keep the vertical
offset level indicator stationary.
Fixed offset
: when the vertical gain is changed, the oscilloscope will have the vertical
offset level indicator move with the actual voltage level.
Press the
Horizontal
softkey to select
Fixed Delay
or
Fixed Position
.
Fixed Position
: when the timebase is changed, the oscilloscope will keep the horizontal
offset indicator stationary.
Fixed Delay
: when the timebase is changed, the oscilloscope will have the horizontal
offset indicator move with the trigger point.
Power On Line
Press the
Utility
button on the front panel; then; press
Power On Line
softkey to select
Enable
or
Disable.
Enable:
when the oscilloscope is energized, the oscilloscope will start automatically
by itself.
Disable
: when the oscilloscope is energized; then; press the
power
key at the
lower-left corner of the front panel to start the oscilloscope.