SDS1000X-E User Manual
8
WWW.SIGLENT.COM
follows.
1. Set the switch to 10X on the probe.
2. Perform steps 1, 2, 3 and 4 of “
Function Inspection
” in the previous section.
3. Check the waveforms displayed and compare them with the following:
Under Perfectly Over
Compensated Compensated Compensated
4. Use a nonmetallic driver to adjust the low-frequency compensation adjustment hole
on the probe until the waveform displayed is as the “Perfectly compensated” in the
figure above.