SDS1000X-E User Manual
46
WWW.SIGLENT.COM
Acquisition Mode
The acquisition mode is used to control how to generate waveform points from sample
points. The oscilloscope provides the following acquisition mode: Normal, Peak Detect,
Average and Eres.
1. Press the
Acquire
button on the front panel to enter the ACQUIRE function menu;
2. Press the
Acquisition
softkey; then turn the
Universal Knob
to select the desired
acquisition mode and push down the knob to confirm. The default setup is
Normal
.
Normal
In this mode, the oscilloscope samples the signal at equal time interval to rebuild the
waveform. For most of the waveforms, the best display effect can be obtained using this
mode. It is the default acquisition mode.
Figure 19 Acquisition System
Peak Detect
In this mode, the oscilloscope acquires the maximum and minimum values of the signal