SDS1000X-E User Manual
158
WWW.SIGLENT.COM
Note:
There are two indicators on the side of the SLA1016, one is red (Power), another
is green (Ready)
The red indicator: If SLA1016s power is normal, this indicator turns on, or
else it blinks or turns off.
The green indicator: When the SLA1016 is connect to the oscilloscope, this
indicator turns on for seconds then turns off, and then blinks, and finally keeps
the light on. When data is transferred between the SLA1016 and the
oscilloscope, the indicator blinks.
Connect the Digital Probes
1. If necessary, turn off the power supply to the device-under-test.
Turning off the power to the device-under-test prevents damage that might occur
if you accidentally short lines together while connecting probes. You can leave
the oscilloscope powered on because no voltage appears at the probes.
2. Connect one of the digital probe cable ports to the digital channels connector on the
front panel of the SDS1000X-E, and the other one to the digital channels connector of
the digital probe.
The digital probe cable is keyed so you can connect it only one way. You do not
need to power-off the oscilloscope.
3. Connect a fly line to one of the digital probe ground pins, connect a grabber to the fly
line, and then connect the grabber to a ground trace/pin from the device-under-test.
The ground lead improves signal fidelity to the oscilloscope, ensuring accurate
measurements,
4. Connect a fly line to one of the digital probe digital channel pins; connect a grabber to
the fly line, and then connect the grabber to a node in the circuit you want to test.
5. Repeat step 4 until you have connected all points of interest.