SDS1000X-E User Manual
130
WWW.SIGLENT.COM
Make Cursor Measurements
1. Press the
Cursors
button on the front panel to enter the CURSOR function menu.
2. Press the
Mode
softkey and set the cursors mode to
Manual
or
Track
.
3. Press the
Source
softkey, and then use the
Universal Knob
to select the desired
source. Only analog channels, math waveforms and reference waveforms that are
displayed are available for cursors.
4. Press the
X Ref
and
Y Ref
softkey to set the reference of X cursors and Y cursors.
Position
: when the horizontal/vertical scale is changed, the X/Y cursors remain
fixed to the grid position on the display.
Delay/Offset
: when the horizontal/vertical is changed, the value of X/Y cursors
remain fixed.
5. To make cursor measurements:
To measure the horizontal time, use the
Universal Knob
to move the X1 and X2
cursors to desired place. If necessary, set the cursor type to
X2-X1
, move X1 and
X2 cursors together.
To measure vertical voltage or current, use the
Universal Knob
to move the Y1
and Y2 cursors to desired place. If necessary, set the cursor type to Y2-Y1,
move Y1 and Y2 cursors together.
To adjust the transparence of the cursors message box, press the
Display/Persist
softkey and go to the second page, press the
Transparence
(20% to 80%) softkey and then turn the
Universal Knob
to adjust the
transparence to the desired value.
Cursor examples:
Use cursors to measure pulse width: