SDS1000X-E User Manual
178
WWW.SIGLENT.COM
state is set to ON, overvoltage protection of AWG output port take effect once the
condition is meeting (the absolute value of input voltage is higher than 4V±0.5V). When
overvoltage protection occurs, a message will be displayed and the output is disabled.
Default
Press
Utility
NextPage4/4
AWG
Setting
Default
to recover the default values of
AWG (Sine, 1.000 KHz, 4.000Vpp, 0.0mVdc, High-Z, Sync output off, OVP On).
Zero Adjust
Press
Utility
NextPage4/4
AWG
Setting
Zero
Adjust
. The AWG can be calibrated
by using
Auto
mode or
Manual
mode. The adoption standard of AWG calibrated is: when
AWG outputs 0V DC signal, the corresponding channel at 1mv/div gear' mean value of
oscilloscope is rang in
±1mv.
Auto mode:
This mode can only be performed with the channel1 of oscilloscope.
Connect the output of AWG to the oscilloscopechannel1, press the Auto soft key,
then the AWG starts to automatic calibrate, and displaysZero Adjust Success!”.
Manual mode:
The AWG can be manually calibrated through any channel on the
oscilloscope.
Take channel 2 as an example:
a. Connect the output of AWG to the oscilloscope channel 2, open the channel 2,
and set DC coupling, open bandwidth limit, Probe is 1X.
b. Adjust the voltage of channel 2 channel to 1mv/div, and make the waveform
visible in the oscilloscope interface, press the
Measure
and measure the mean
value of channel 2.
c. Press the
Manua
l and rotate the knob to adjust the waveform, so that the mean
value of the observed CH2 will be in the range of ±1mv, and then press the
Save
,
the oscilloscope displays "Zero Adjust Success!".
Systems
System Status
Press
Utility
NextPage4/4
AWG
System
, then select the System Status option to view
the AWG’s system information, including software version, hardware version, and serial
number. After finished, press the
return
softkey to exit the system information page.