SDS1000X-E User Manual
75
WWW.SIGLENT.COM
level.
Figure 32 Absolute Window Trigger
To set window trigger via Relative window type:
1. Press the
Setup
button on the front panel to enter the TRIGGER system function
menu.
2. Press the
Type
softkey; then use the
Universal Knob
to select
Window
and push
down the knob to confirm.
3. Press the
Source
softkey; turn the
Universal Knob
to select CH1 or CH2 as the trigger
source.
4. Press the
Window Type
softkey to select
Relative
.
5. Press the
Center
Delta
softkey to select
Center
or
Delta
trigger level mode; then turn
the
Trigger Level Knob
to adjust the position. The
Center
and
Delta
values are
displayed at the upper right corner of the screen.
6. In the trigger state message box,
C
means
Center
, the center value of the lower and
upper trigger levels;
D
means
Delta
, the difference between the lower (or upper)
trigger level and the trigger level center.