SDS1000X-E User Manual
113
WWW.SIGLENT.COM
Note: The vertical Variable knob and the vertical Position knob are only works for
the reference waveform and the math waveform. Please make a difference with the
channel vertical system.
Adjust REF Waveform
1. Please refer to the
To Display REF Waveform
above to display the desired
reference waveform.
2. Press the
Scale
and
Position
softkey and turn the
Universal Knob
to adjust the
vertical scale and position of the reference waveform. The vertical scale and position
information display at the middle of the screen.
The initial values display at the middle of the screen is the setup that when the reference
waveform been saved.
Figure 67 Reference Waveform
Clear REF Waveform
The oscilloscope does not have the Clear option under the REF WAVE function menu.
To clear the appointed reference waveform, you can save a new reference waveform to