SDS1000X-E User Manual
197
WWW.SIGLENT.COM
Example
The following are examples to familiarize you with bode plot.
Simple Sweep Operation
Vari-level Sweep Operation
Simple Sweep Operation
1. Press
Utility
button on the front panel to enter the UTILITY function menu, and then
press the
Next Page
soft key to enter the second page. Press
Bode Plot Ⅱ
to enter
the BODE PLOT function menu. Press
Config
to enter Configure menu.
2. Press
Source
to enter Source menu to select the waveform generator. In this
example, press
Interface
to select “
USB
”, and then press
Connection Test
to verify
setting. Then return to the Configure menu.
Figure 133 Bode Plot Interface Connection
3. Press
Sweep Mode
to select “
Simple
”.
Figure 134 Sweep Configure Menu
4. Press
Set Sweep
to enter Set Sweep menu. In this example, press
Mode
to select
“
Decade
” to sweep frequency logarithmically. Press
Start
and use the
Universal
Knob
to set the start frequency to 100 Hz. Then set stop frequency to 1 MHz. Press
Points/dec
and use the
Universal Knob
to set the frequency point to 20 points per
decade. Then return to the Configure menu.
Figure 135 Set Sweep Parameters Menu
5. Press
Set Stimulus
to enter Set Stimulus menu. Press
Amplitude
and use the
Universal Knob
to set the signal amplitude to 2 V. All the other options remain their
default setting. Then return to the Configure menu.