Instrument Functions
R&S
®
NGP800
94User Manual 5601.5610.02 ─ 04
Select "Yes" to restore factory adjustment.
3. To proceed channel adjustment, select "User Adjustment" in Figure 6-46.
The R&S NGP800 displays the "ADJUSTMENT" wizard to guide the channel
adjustment procedures.
Figure 6-47: Channel adjustment wizard
4. Depending on the types of adjustment (voltage or current), setup the instruments
illustrated in
Figure 6-44 or Figure 6-45.
5. Follow the on-screen instructions displayed in the Figure 6-47.
The R&S NGP800 applies a low voltage/current followed by a high voltage/current
for voltage/current adjustment. Key in the measured value from DMM using the on-
screen keypad. See
Figure 6-48.
Figure 6-48: Channel adjustment procedure
6.
Confirm the entry with
.
7. Leave the setup connections as open.
Select "Proceed" to start the voltage adjustment automatically.
8. Follow the on-screen instructions for current adjustment.