Remote Control Commands
R&S
®
NGP800
105User Manual 5601.5610.02 ─ 04
SYSTem:UPTime?
Queries system uptime.
Usage: Query only
7.3 Display Commands
The DISPlay subsystem contains the commands for display functions, which do not
affect signal generation directly.
DISPlay:BRIGhtness...................................................................................................... 105
DISPlay[:WINDow]:TEXT:CLEar...................................................................................... 105
DISPlay[:WINDow]:TEXT[:DATA]..................................................................................... 105
DISPlay:BRIGhtness <brightness>
DISPlay:BRIGhtness?
Sets or queries the display brightness.
Parameters:
<brightness> Displays brightness for the instrument.
Range: 0.0 to 1.0
Increment: 0.1
*RST: 0.8
Example:
DISPlay:BRIGhtness 0.5
DISPlay:BRIGhtness? -> 0.5
Returns the display brightness value.
DISPlay[:WINDow]:TEXT:CLEar
Clears the text message box on the front display.
Usage: Event
DISPlay[:WINDow]:TEXT[:DATA] <text>
Shows the text message box on the front display.
Setting parameters:
<text> New value for text message box.
Usage: Setting only
7.4 Trigger Commands
The TRIGger subsystem contains the commands for DIO signal triggering.
Trigger Commands
Loading ...
Loading ...